Applications
USP Alosetron Organic Impurities Sensitivity Solution on Gemini 3 μm C18 75 x 4.6 mm using Agilent 1260 Binary UHPLC
26357
Separation Mode: Reversed Phase
Reversed Phase
USP
C18
HPLC
Serotonin Antagonists
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

USP Alosetron Organic Impurities Sensitivity Solution on Gemini 3 μm C18 75 x 4.6 mm using Agilent 1260 Binary UHPLC

Analytes

1

Analyte not found

Conditions

LC Conditions (App ID: 26357)
Official Method:USP
Column:Gemini 3 µm C18 110 Å, LC Column 75 x 4.6 mm, Ea
Brand Name:Gemini
Part No:00C-4439-E0
Elution Type:Gradient
Mobile Phase:A: 0.02 M monobasic sodium phosphate. Adjust with phosphoric acid to a pH of 3.0
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
109010
259010
3205050
4255050
5269010
6309010
Flow Rate:1 mL/min
Temperature:30°C
System:Agilent Technologies Agilent Technologies
Detection:LC/UV (DAD, PDA) nm (nanometers) @ 0°C