USP Butamben Impurity Standard Solution 0.2 µg/mL on Kinetex 5 µm XB-C18 100 x 2.1 mm on Agilent 1260 UHPLC
26338
Separation Mode: Reversed Phase
Reversed Phase
USP
XB-C18
HPLC
Anesthetics, Local
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

USP Butamben Impurity Standard Solution 0.2 µg/mL on Kinetex 5 µm XB-C18 100 x 2.1 mm on Agilent 1260 UHPLC
Analytes
1
2
Compound Name
4-Aminobenzoic Acid
Details
Official Method: USP
LC Conditions (App ID: 26338)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
109010
219010
365050
48.55050
58.69010
6109010
Flow Rate:
0.4 mL/min
Temperature:
30°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/UV (DAD, PDA) nm (nanometers) @ 0°C