Lisinopril Ph. Eur. Test Solution on Kinetex 5 µm C18 250 x 4.6 mm on Agilent 1100 LC/UV
26297
Separation Mode: Reversed Phase
Reversed Phase
Ph.Eur. M.1120
C18
HPLC
Antihypertensive Agents
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Lisinopril Ph. Eur. Test Solution on Kinetex 5 µm C18 250 x 4.6 mm on Agilent 1100 LC/UV
Analytes
1
2
3
4

Analyte not found

Details
Official Method: Ph.Eur. M.1120
LC Conditions (App ID: 26297)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: Mix 3 volumes of acetonitrile R1 and 97 volumes of 3.12 g/L solution of sodium dihydrogen phosphate R previously adjusted to pH 3.8 with dilute phosphoric acid R
B: Mix 20.5 volumes of acetonitrile R1 and 79.5 volumes of 3.12 g/L solution of phosphoric acid R
Gradient Profile:
Step No.Time(min)%A%B
10100
22100
3370100
4620100
562.5100
666100
Flow Rate:
1.6 mL/min
Temperature:
50°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/UV (DAD, PDA) nm (nanometers) @ 0°C