Applications
Dexamethasone USP Standard Solution on Kinetex 1.7 µm C18 100 x 2.1 mm on Agilent 1290 UHPLC
26286
Separation Mode: Reversed Phase
Reversed Phase
USP
C18
HPLC
Antineoplastic Agents
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Dexamethasone USP Standard Solution on Kinetex 1.7 µm C18 100 x 2.1 mm on Agilent 1290 UHPLC
Analytes
1
Compound Name
Dexamethasone
Details
Official Method: USP
LC Conditions (App ID: 26286)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 3.4 g/L of monobasic potassium phosphate in water. Adjust with phosphoric acid to a pH of 3.0
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
107624
2107624
3154555
4161090
516.17624
6207624
Flow Rate:
0.4 mL/min
Temperature:
35°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/UV (DAD, PDA) nm (nanometers) @ 0°C

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Gemini 3 µm C18 110 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase