Applications
Rapid Sample Extraction of Chloroquine, Hydroxychloroquine, and Desethylchloroquine from Serum for LC-MS/MS Analysis
25942
Separation Mode: Reversed Phase
Reversed Phase
F5
HPLC
Antirheumatic Agents
Clinical
Pharmaceutical/Biopharmaceutical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Rapid Sample Extraction of Chloroquine, Hydroxychloroquine, and Desethylchloroquine from Serum for LC-MS/MS Analysis

Analytes

1
2
3

Analyte not found

Conditions

LC Conditions (App ID: 25942)
Column:Kinetex 2.6 µm F5 100 Å, LC Column 50 x 2.1 mm, Ea
Brand Name:Kinetex
Part No:00B-4723-AN
Elution Type:Gradient
Mobile Phase:A: 0.1 % Formic acid in water
B: Methanol
Gradient Profile:
Step No.Time(min)%A%B
10982
22937
338515
43.26040
54.2595
66982
Flow Rate:0.5 mL/min
Temperature:20°C
System:Agilent Technologies Agilent Technologies
Detection:LC/MS/MS @ 0°C

Sample Preparation Details

Order Items Used in This Application

Similar Applications
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase