Applications
Lopinavir & Ritonavir Tablets USP (COVID-19) Kinetex 5 µm C8 150 x 4.6 mm
25916
Separation Mode: Reversed Phase
Reversed Phase
USP
C8
HPLC
Anti-HIV Agents
Pharmaceutical/Biopharmaceutical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Lopinavir & Ritonavir Tablets USP (COVID-19) Kinetex 5 µm C8 150 x 4.6 mm

Analytes

1
2

Analyte not found

Conditions

LC Conditions (App ID: 25916)
Official Method:USP
Column:Kinetex 5 µm C8 100 Å, LC Column 150 x 4.6 mm, Ea
Brand Name:Kinetex
Part No:00F-4608-E0
Elution Type:Gradient
Mobile Phase:A: (175:100:100:625) of ACN:MeOH:THF:Buffer (4.1 g/L of monobasic potassium phosphate in water)
Gradient Profile:
Step No.Time(min)%A
10100
250100
Flow Rate:1.5 mL/min
Temperature:40°C
System:Agilent Technologies 1260 Infinity Hybrid SFC/UHPLC Agilent Technologies 1260 Infinity Hybrid SFC/UHPLC
Detection:LC/UV (DAD, PDA) nm (nanometers) @ 0°C