Applications
Clean-Up of Phospholipids in Human Plasma Comparing PPT VS. Strata-X PRO SPE on a Kinetex 2.6 µm C18 50 x 2.1 mm
25779
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Clinical
Clinical Research
Therapeutic / Clinical

Applications

Clean-Up of Phospholipids in Human Plasma Comparing PPT VS. Strata-X PRO SPE on a Kinetex 2.6 µm C18 50 x 2.1 mm
Analytes
1
2
3

Analyte not found

Details
LC Conditions (App ID: 25779)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid in water
B: 0.1% Formic acid in methanol
Gradient Profile:
Step No.Time(min)%A%B
106040
20.5595
311.5595
411.516040
513.56040
Flow Rate:
400 μL/min
Temperature:
40°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/MS/MS @ 40°C
Sample Preparation Details
Description:
Strata-X PRO, 96-Well Plate, 30mg/well, Ea
Part Number:
Pre Treatment Note:
Sample Pre-treatment: • Add 100 µL Serum to tube • Add 900 µL of chilled (~0°C) ACN to the serum while vortexing • Centrifuge samples at 6000 rpm for 5 minutes and transfer supernatant

Order Items Used in This Application

Similar Applications
search
No Data