Applications
Pain Panel from Urine using Strata DE SLE and a Kinetex Biphenyl Core-Shell LC Column
24555
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Analgesics, Opioid
Clinical
Forensics/Toxicology

Applications

Pain Panel from Urine using Strata DE SLE and a Kinetex Biphenyl Core-Shell LC Column
Analytes
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Compound Name
Morphine
Details
LC Conditions (App ID: 24555)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 0.1% formic acid in water
B: 0.1% formic acid in acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10955
240100
350100
45.01955
57.1955
Flow Rate:
500 µL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) @ 22°C

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm HILIC 100 Å, LC Column 50 x 4.6 mm, Ea
Hydrophilic Interaction
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA
Improved Performance
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm HILIC 100 Å, LC Column 50 x 4.6 mm, Ea
Hydrophilic Interaction
Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA
Improved Performance