Applications
Low levels of B12 analysis from Vitamin pill formulation on Kinetex 5u EVO 150x4.6mm
22916
Separation Mode: Reversed Phase
Reversed Phase
EVO C18
HPLC
Vitamin B Complex
Clinical
Clinical Research
Therapeutic / Clinical

Applications

Low levels of B12 analysis from Vitamin pill formulation on Kinetex 5u EVO 150x4.6mm
Analytes
1
2
3
4
Compound Name
Hexa-Betalin
Details
LC Conditions (App ID: 22916)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 3g Sodium 1-Hexanesulfonate Monohydrate with 0.4mL TEA and 15mL Acetic Acid with 1985mL Water
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10955
210298
312298
412.1955
515955
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) nm @ 22°C

Order Items Used in This Application

Similar Applications
search