Applications
Mycotoxins extracted from corn using roQ by LC/MS/MS (Positive mode)
22711
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Antineoplastic Agents

Applications

Mycotoxins extracted from corn using roQ by LC/MS/MS (Positive mode)

Analytes

1
2
3
4
5
6
7
8
9
10
11
12

Analyte not found

Conditions

LC Conditions (App ID: 22711)
Column:Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand Name:Kinetex
Part No:00B-4496-AN
Elution Type:Gradient
Mobile Phase:A: 5 mM ammonium acetate w/ 0.5% acetic acid
B: 5 mM ammonium acetate in methanol w/ 0.5% acetic acid
Gradient Profile:
Step No.Time(min)%A%B
10955
22955
352080
45.2298
58298
Flow Rate:0.45 mL/min
Temperature:22°C
Detection:Tandem Mass Spec (MS-MS) @ 22°C

Order Items Used in This Application

Similar Applications
No Data