Applications
Carbamate pesticides on Kinetex C18, 5u with LC/MS/MS detection
22064
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Molluscacides
Environmental
Agriculture
Product / Trade Good
Environmental / Health
Environmental / Health

Applications

Carbamate pesticides on Kinetex C18, 5u with LC/MS/MS detection

Analytes

1
2
3
4
5
6
7
8
Compound Name
Dioxamyl
CID
31657
Molecular Formula
C7H13N3O3S
Molecular Weight
219
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1

Conditions

LC Conditions (App ID: 22064)
Column:Kinetex 5 µm C18 100 Å, LC Column 150 x 2.1 mm, Ea
Brand Name:Kinetex
Part No:00F-4601-AN
Elution Type:Gradient
Mobile Phase:A: 0.1% formic acid in DI H2O
B: 0.1% formic acid in Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
109010
2203070
3223070
422.19010
5279010
Flow Rate:200 µL/min
Temperature:42°C
Detection:Mass Spectrometer (MS) amu @ 0°C

Order Items Used in This Application

Similar Applications
Kinetex 2.6 µm Polar C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
EnviroSep-PP 5 µm PP, LC Column 125 x 4.6 mm, Ea
Reversed Phase