Applications
Aldosterone analysis by LCMS using Kinetex 2.6um XB-C18 50x3.0mm
20648
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Clinical
Clinical Research
Therapeutic / Clinical

Applications

Aldosterone analysis by LCMS using Kinetex 2.6um XB-C18 50x3.0mm
Analytes
1
Compound Name
Aldosterone
Details
LC Conditions (App ID: 20648)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: Water
B: 50/50 AcCN/Methanol
Gradient Profile:
Step No.Time(min)%A%B
107030
241090
351090
45.17030
577030
Flow Rate:
0.5 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) @ 22°C
Sample Preparation Details
Description:
Strata-X-A 33 µm Polymeric Strong Anion, 60 mg / 3 mL, Tubes, 50/Pk
Part Number:
SPE Condition 1:
3 mL Methanol
SPE Condition 2:
3 mL 25 mM Ammonium Bicarbonate
SPE Load:
Load 0.50 mL Sample diluted in 25 mM Ammonium Bicarbonate
SPE Wash 1:
3 mL 25mm ammonium bicarbonate buffer
SPE Wash 2:
3 mL 25% methanol in water
SPE Dry:
5 min at high vacuum
SPE Elution 1:
1 mL Ethyl Acetate:Isopropanol:NH4OH (7:2:1)
SPE Elution 2:
1 mL Ethyl Acetate:Isopropanol:NH4OH (7:2:1)
SPE Evaporation:
Evaporate to dryness
SPE Reconstitute:
0 µL initial mobile phase

Order Items Used in This Application

Similar Applications