Applications
LC/MS/MS Analysis of Digoxin and Digitoxin in plasma (0.25 ng/mL) using Strata-X and Kinetex C8
19691
Separation Mode: Reversed Phase
Reversed Phase
C8
HPLC
Anti-Arrhythmia Agents
Clinical

Applications

LC/MS/MS Analysis of Digoxin and Digitoxin in plasma (0.25 ng/mL) using Strata-X and Kinetex C8
Analytes
1
2
3
Compound Name
Digoxin
Details
LC Conditions (App ID: 19691)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 10mM Ammonium acetate
B: 10mM Ammonium acetate in methanol
Gradient Profile:
Step No.Time(min)%A%B
105050
22.50100
32.515050
455050
Flow Rate:
400 µL/min
Temperature:
30°C
Detection:
Mass Spectrometer (MS) @ 350°C
Sample Preparation Details
Description:
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:
SPE Condition 1:
2 mL Methanol at 2 mL/min
SPE Condition 2:
2 mL 10mM Ammonium acetate at 2 mL/min
SPE Load:
Load 3.00 mL Sample diluted in Water
SPE Wash 1:
1 mL 50:50 10mM ammonium acetate/10mM ammonium acetate in methanol at 2 mL/min
SPE Dry:
10min at full vacuum
SPE Elution 1:
2 mL Methanol at 1 mL/min
SPE Evaporation:
Evaporate to dryness
SPE Reconstitute:
100 µL 50:50 Methanol/water