Applications
Cyanuric acid and Melamine on Kinetex HILIC 2.6µm 50x2.1mm by LCMS
19170
Separation Mode: Hydrophilic Interaction
Hydrophilic Interaction
HILIC
HPLC
Food & Beverage

Applications

Cyanuric acid and Melamine on Kinetex HILIC 2.6µm 50x2.1mm by LCMS
Analytes
1
2
3
4
Compound Name
Isocyanuric Acid
Details
LC Conditions (App ID: 19170)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 50/40/10 ACN/H2O/100mM ammonium acetate pH5.8
B: 90/10 ACN/100mM ammonium acetate pH5.8
Gradient Profile:
Step No.Time(min)%A%B
100100
20.80100
30.811000
41.11000
51.110100
63.80100
Flow Rate:
700 µL/min
Temperature:
25°C
System:
Agilent (HP) 1100
Detection:
Mass Spectrometer (MS) amu @ 22°C