Applications
Extraction of Expanded Amphetamines Using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
18847
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Serotonin Agents
Forensics/Toxicology
Clinical

Applications

Extraction of Expanded Amphetamines Using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS

Analytes

1
2
3
4
5
6
7
8
9
10
11
Compound Name
Hmma
CID
3081137
Molecular Formula
C11H17NO2
Molecular Weight
195
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
2

Conditions

LC Conditions (App ID: 18847)
Column:Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand Name:Kinetex
Part No:00B-4462-AN
Elution Type:Gradient
Mobile Phase:A: 0.1%FA-5mM ammonium acetate in H2O
B: 0.1%FA-5mM ammonium acetate in Methanol
Gradient Profile:
Step No.Time(min)%A%B
109010
236040
33.02595
43.3595
53.319010
65.19010
Flow Rate:450 µL/min
Temperature:25°C
Detection:Mass Spectrometer (MS) amu @ 22°C

Sample Preparation Details
Description:Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:8B-S029-TBJ
Pre Treatment Note:Blown down samples (after evaporation) were reconstituted in 1 mL of initial mobile phase (A/B=90/10).
SPE Condition 1:1 mL Methanol at 2 mL/min
SPE Condition 2:1 mL Water at 2 mL/min
SPE Load:Load 1.00 mL Sample diluted in 100 mM KH2PO4 (pH 6)
SPE Wash 1:1 mL 0.1 N HCL at 2 mL/min
SPE Wash 2:1 mL Methanol at 2 mL/min
SPE Dry:2 minutes at 10" of Hg pressure
SPE Elution 1:0.5 mL 5% Ammonium hydroxide/methanol at 1 mL/min
SPE Elution 2:0.5 mL 5% Ammonium hydroxide/methanol at 1 mL/min
SPE Evaporation:Evaporate to dryness
SPE Reconstitute:1000 µL Initial mobile phase (A/B=90/10)

Order Items Used in This Application

Similar Applications