Applications
Extraction of 6-MAM using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
18665
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Forensics/Toxicology

Applications

Extraction of 6-MAM using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS

Analytes

1
2
Compound Name
6-Acetylmorphine
CID
5462507
Molecular Formula
C19H21NO4
Molecular Weight
327
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
1

Conditions

LC Conditions (App ID: 18665)
Column:Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand Name:Kinetex
Part No:00B-4462-AN
Elution Type:Gradient
Mobile Phase:A: 0.1% Formic acid in H2O
B: 0.1% Formic acid in ACN
Gradient Profile:
Step No.Time(min)%A%B
10973
20.24973
33.02595
43.3595
53.31973
64.97973
Flow Rate:450 µL/min
Temperature:25°C
Detection:Mass Spectrometer (MS) amu @ 22°C

Sample Preparation Details
Description:Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:8B-S029-TBJ
SPE Condition 1:1 mL Methanol at 2 mL/min
SPE Condition 2:1 mL Water at 2 mL/min
SPE Load:Load 1.00 mL Sample diluted in 1
SPE Wash 1:1 mL 0.1N HCl at 2 mL/min
SPE Wash 2:1 mL Methanol at 2 mL/min
SPE Dry:2-3 mins at 10" of Hg
SPE Elution 1:0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min
SPE Elution 2:0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min
SPE Evaporation:Evaporate to dryness
SPE Reconstitute:200 µL Initial LC mobile phase (A/B=97/3)

Order Items Used in This Application

Similar Applications