Applications
EPA 531.1 Carbamate Pesticides at High Speed using Kinetex 2.6µm C18 50 x 2.1mm ID
18610
Separation Mode: Reversed Phase
Reversed Phase
EPA 531
C18
HPLC
Antinematodal Agents

Applications

EPA 531.1 Carbamate Pesticides at High Speed using Kinetex 2.6µm C18 50 x 2.1mm ID
Analytes
1
2
3
4
5
6
7
8
9
10
11
12
Compound Name
Temik Sulfoxide
Details
Official Method: EPA 531
LC Conditions (App ID: 18610)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% H3PO4
B: Acetonitrile with 0.1% H3PO4
Gradient Profile:
Step No.Time(min)%A%B
10955
23595
Flow Rate:
1 mL/min
Temperature:
40°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) nm @ 22°C
Sample Preparation Details
Description:
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:
SPE Condition 1:
1 mL Methanol
SPE Condition 2:
1 mL Loading buffer
SPE Load:
Load Sample
SPE Wash 1:
1 mL Loading buffer
SPE Wash 2:
1 mL Methanol/Water (15:85)
SPE Dry:
2 minutes at 10 mm Hg vacuum
SPE Elution 1:
0.5 mL 5% Formic acid in Methanol
SPE Elution 2:
0.5 mL 5% Formic acid in Methanol
SPE Evaporation:
Evaporate to dryness
SPE Reconstitute:
1 mL Mobile phase

Order Items Used in This Application

Similar Applications
search
No Data