Extraction of Barbiturates from Urine on Strata™ X
14622
Separation Mode: Reversed Phase
Reversed Phase
Max-RP
HPLC
Adjuvants, Anesthesia
Forensics/Toxicology
Pharmaceutical
Therapeutic / Clinical

Applications

Extraction of Barbiturates from Urine on Strata™ X
Analytes
1
2
3
4
5
Compound Name
Phenobarbital
Details
LC Conditions (App ID: 14622)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 20 mM KH2PO4, pH 7.0
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
108020
238020
3155050
4168020
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Fixed Wave. (UV) nm @ 22°C
Sample Preparation Details
Description:
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part Number:
Pre Treatment Note:
Blow down under slow stream of N2 @ 40° C and reconstitute with 300 µL water
SPE Condition 1:
1.0 mL Methanol
SPE Condition 2:
1.0 mL Water
SPE Load:
Load 1.00 mL Sample diluted in KH2PO4 (100mM) pH 6.0
SPE Wash 1:
1.0 mL Methanol / Water (5:95)
SPE Wash 2:
1.0 mL Methanol / 2% CH3COOH in Water (10:90)
SPE Elution 1:
1.0 mL Methanol / 2% Ammonium Hydroxide in Water (35:65)
SPE Evaporation:
Evaporate to dryness
SPE Reconstitute:
300 µL Water

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm EVO C18 100 Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-5MSi, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Standard