Zebron ZB-1701P is EPA Method 8081 certified GC column
Zebron ZB-1701P GC Columns
Enhanced Response for DDT and Endrin

  • Specially tested to ensure response of DDT, Endrin, Endrin Aldehyde, and Endrin Ketone
  • Guaranteed column for pesticide analysis
  • EPA Method 8081 Certified

Want an even more rugged alternative for pesticide analysis? See ZB-MutliResidue-1

Recommended Use
Aroclors, nitrogen containing pesticides, organochlorine pesticides, organophosphorous pesticides

zb1701P

Phase
ZB-1701P
Temperature Limits
-20 to 280/300
Recommended Use
Guaranteed for pesticide analysis and certified for EPA Method 8081
Upgrade to Zebron from any 14 % cyanopropylphenyl / 86 % dimethylpolysiloxane phase:

Agilent® Restek® SGE® Supelco® OV®
DB®-1701 Rtx®-1701 BP10 SPB®-1701 OV-1701
DB-1701P Rtx-VMS   Equity®-1701  
CP-Sil 19 CB        

Chlorinated Pesticides by GC-ECD: EPA Method 8081

PAHs and PCBs In A Single Run

Column:
Zebron ZB-1701P
Dimension:
30 meter x 0.25 mm x 0.25 µm
Part No.:
Injection Volume:
Splitless @ 240 °C, 1 µL
Carrier Gas:
Helium @ 1.6 mL/min (constant flow)
Oven Program:
100 °C to 200 °C @ 25 °C/min to 240 °C @ 6 °C/min to 265 °C @ 20 °C/min for 5 min
Detector:
ECD @ 300 °C

Sample:
All compounds are 20 ppm
1. 1-Bromo-2-Nitrobenzene (IS)
2. Tetrachloro-m-xylene (Surr)
3. α-BHC
4. Pentachloronitrobenzene (IS)
5. γ-BHC (Lindane)
6. Heptachlor
7. Aldrin
8. β-BHC
9. δ-BHC
10. Heptachlor Epoxide
11. Endosulfan I
Sample:
12. γ-Chlordane
13. α-Chlordane
14. DDE
15. Dieldrin
16. Endrin
17. DDD
18. Endosulfan II
19. DDT
20. Endrin Aldehyde
21. Methoxychlor
22. Endosulfan Sulfate
23. Endrin Ketone
24. Decachlorobiphenyl (Surr)